7ML19985GE01 SITRANS LC 500 – INSTRUCTION MANUAL Page 35
mmmmm
Interconnection
• The Upper Sensor Limit (USL) and Lower Sensor Limit (LSL) are set to 3300 and
1.666 pF respectively, and the following conditions apply: the Upper Range Value
(URV) and Lower Range Value (LRV) should be within the USL to LSL range, but
can be set anywhere within that range.
• An interruption of the measuring connection will be detected: a loose or
interrupted connection results in up to 0.5 pF capacitance, which is below the
adjusted LSL and thus signals a FAULT condition.
DATE "customer input data via HART"
SENSOR SERIAL NUMBER "customer input data via HART"
FINAL ASSEMBLY NUMBER "customer input data via HART"
TV1 Units UNDEFINED
TV1 LRV 0 [switch position (0)E, TV1]
TV! URV 1.0 [switch position (0)F, TV1]
Setting Description