Chapter 2 Command Summary
HP 53131A/132A Command Summary
2-30 Programming Guide
Table 2-2. HP 53131A/132A SCPI Command Summary (Continued)
Keyword/Syntax Parameter Form Std/
New
Comments
Measurement Instructions*
:CONFigure[:SCALar]:<function>
:CONFigure?
:MEASure[:SCALar]:<function>?
:READ[[:SCALar]:<function>]?
:FETCh[[:SCALar]:<function>]?
See <parameters> and
<source_list> in table
on the next page.
See <parameters> and
<source_list> in table
on the next page.
Std
Std
Std
Std
Std
Configures instrument to perform
specified measurement.
Returns function configured by the last
:CONF or :MEAS command.
Configures instrument, initiates
measurement, and queries for the result
(i.e., provides complete measurement
sequence).
Initiates measurement, and queries for the
result. (Performs a :FETCh? on fresh
data.)
Queries the result.
*The <function> and corresponding <parameters> and <source list> are defined by the following listing in this table
(see next page).