HP (Hewlett-Packard) waterdispenser 132A Handleiding

PDF Handleidingen
330 Pagina's Nederlands

HP (Hewlett-Packard) waterdispenser 132A
Chapter 4 Command Reference
*ESR?
(Event Status Register Query)
4-128 Programming Guide
*ESR?
(Event Status Register Query)
*ESR?
Queries the Standard Event Status Register, shown in Figure 4-5.
This event register captures changes in conditions, by having each event bit
correspond to a specific condition in the instrument.
An event becomes TRUE when the associated condition makes the defined
transition. The event bits, once set, are sticky.That is,
they cannot be cleared even if they do not reflect the current status of a related
condition, until they are read.
This register is cleared by *CLS, by *ESR?, and at power-on.
Note that the instruments power-on sequence initially clears the register, but then
records any subsequent events during the power-on sequence including setting the
PON (power on) bit.
Figure 4-5. Standard Event Status Register
See the section titled “Standard Event Status Register Group,
page 3-24, in Chapter 3 of this guide for a detailed description of the Standard
Event Status Register.
Numeric data transferred as ASCII bytes in <NR1> format.
Range is 0 to 255.
The query response is an integer formed by the binary-weighting of the bits.
The value of unused bit is zero.
Query Response
Gerelateerde Handleidingen voor HP (Hewlett-Packard) waterdispenser 132A